Intelligent module communication protocol Introduction
Communication port: RS485 Port setting: 9600,N,8,1 (by default) Device address: 0x01 (by default) Protocol specifications: Modbus RTU Commands support: 0x03 read register 0X06 write register| 0x10 continuous write register Information frame format
0x03 read data [HEX]
01
03
×× ××
×× ××
×× ××
Address
Function code
Data head address
Data length
Check code
0x06 write data [HEX]
01
06
×× ××
×× ××
×× ××
Address
Function code
Data address
Write data
Check code
Remarks: The check code is 16CRC with low byte ahead.
0x10 Continuous write data [HEX]
01
10
×× ××
××××
Address
Function code
Data address
Register number
××
×× ××
×× ××
Byte number
Write data
Check code
Format of register data
Address
Data name
Switch coefficient
Status
0
Temperature
0.1°C
R
1
DO
0.01mg/L
R
2
Saturability
0.1%DO
R
3
Sensor. null point
0.1%
R
4
Sensor. slope
0.1mV
R
5
Sensor. MV
0.1%S
R
6
System status. 01
Format 4*4bit 0xFFFF
R
7
System status.02 User command address
Format: 4*4bit 0xFFFF
R/W
Remarks:Data in each address is a 16-bit signed integer, the length is 2 bytes. The real result=Register data * switch coefficient Status:R=only read; R/W= read/write